Fit assessment of anatomic plates for the distal medial tibia
Schmutz, Beat, Wullschleger, Martin E., Kim, Henry, Noser, Hansrudi, & Schutz, Michael A. (2008) Fit assessment of anatomic plates for the distal medial tibia. Journal of Orthopaedic Trauma, 22(4), pp. 258-263.
Abstract
Objectives: With the development and popularization of minimally invasive surgical methods and implants for fracture fixation, it is increasingly important that the available implants are precontoured to the specific anatomic location for which they are designed. The objective of this study was to develop a noninvasive method and criteria for quantifying the fit of a distal periarticular medial tibia plate and to test the method on a small set of tibia models.
Methods: The undersurface of the plate was extracted from a digital model of the plate. The surface of the plate was fitted to 21 computer tomography (CT)-based 3-dimensional (3-D) models of human tibiae. Four criteria were defined that constitute an anatomic plate fit and subsequently were applied for the quantitative fit assessment. The fitting of the plate undersurface to the bone was entirely conducted in a virtual environment.
Results: An anatomic fit of the plate was achieved for 4 of the models (19%). The individual categories generated fits of 62% (n = 13) for the proximal end; 43% (n = 9) for the proximal angle; 57% (n = 12) for the middle distance; and 57% (n = 12) for a distal fit.
Conclusions: Although for the 4 individual criteria plate fits of 43%-62% were achieved, a global/anatomic fit only occurred for 19% of the bone models. This outcome is likely a result of bone morphology variations, which exist in a random population sample combined with the effects of a nonoptimized plate shape. Recommendations for optimizing the fit of the plate are discussed.
Citations:
Citation countsare sourced monthly from Scopus and Web of Science citation databases.
These databases contain citations from different subsets of available publications and different time periods and thus the citation count from each is usually different. Some works are not in either database and no count is displayed. Scopus includes citations from articles published in 1996 onwards, and Web of Science generally from 1980 onwards.
Citations counts from the Google Scholar™ indexing service can be viewed at the linked Google Scholar™ search.
Full-text downloads:
Full-text downloadsdisplays the total number of times this work’s files (e.g., a PDF) have been downloaded from QUT ePrints as well as the number of downloads in the previous 365 days. The count includes downloads for all files if a work has more than one.
| ID Code: | 13334 |
|---|---|
| Item Type: | Journal Article |
| Keywords: | 3D models, tibia, plate fit, periarticular plates, fracture fixation |
| ISSN: | 1531-2291 |
| Subjects: | Australian and New Zealand Standard Research Classification > ENGINEERING (090000) > BIOMEDICAL ENGINEERING (090300) |
| Divisions: | Past > QUT Faculties & Divisions > Faculty of Built Environment and Engineering Current > Institutes > Institute of Health and Biomedical Innovation |
| Copyright Owner: | Copyright 2008 Lippincott Williams & Wilkins |
| Copyright Statement: | This is the author-version of the work. The final, definitive version of this article has been published in the Journal, Journal of Orthopaedic Trauma 22(4):pp. 258-263, 2008 © < Lippincott, Williams & Wilkins. |
| Deposited On: | 17 Apr 2008 |
| Last Modified: | 29 Feb 2012 23:48 |
Export: EndNote | Dublin Core | BibTeX