Radiotherapy treatment verification using radiological thickness measured with an amorphous silicon electronic portal imaging device: Monte Carlo simulation and experiment
Kairn, Tanya, Cassidy, D., Sandford, Peta M., & Fielding, Andrew L. (2008) Radiotherapy treatment verification using radiological thickness measured with an amorphous silicon electronic portal imaging device: Monte Carlo simulation and experiment. Physics in Medicine and Biology, 53(14), pp. 3903-3919.
Abstract
This work validates the use of an amorphous-silicon, flat-panel electronic portal imaging device (a-Si EPID) for use as a gauge of patient or phantom radiological thickness, as an alternative to dosimetry. The response of the a-Si EPID is calibrated by adapting a technique previously applied to scanning liquid ion chamber EPIDs, and the stability, accuracy and reliability of this calibration are explored in detail. We find that the stability of this calibration, between different linacs at the same centre, is sufficient to justify calibrating only one of the EPIDs every month, and using the calibration data thus obtained to perform measurements on all of the other linacs. Radiological thickness is shown to provide a reliable means of relating experimental measurements to the results of BEAMnrc Monte Carlo simulations of the linac-phantom-EPID system. For these reasons we suggest that radiological thickness can be used to verify radiotherapy treatment delivery and identify changes in the treatment field, patient position and target location, as well as patient physical thickness.
Citations:
Citation countsare sourced monthly from Scopus and Web of Science citation databases.
These databases contain citations from different subsets of available publications and different time periods and thus the citation count from each is usually different. Some works are not in either database and no count is displayed. Scopus includes citations from articles published in 1996 onwards, and Web of Science generally from 1980 onwards.
Citations counts from the Google Scholar™ indexing service can be viewed at the linked Google Scholar™ search.
Full-text downloads:
Full-text downloadsdisplays the total number of times this work’s files (e.g., a PDF) have been downloaded from QUT ePrints as well as the number of downloads in the previous 365 days. The count includes downloads for all files if a work has more than one.
Export: EndNote | Dublin Core | BibTeX
Repository Staff Only: item control page