Calibration of an ingestible temperature sensor
Hunt, Andrew P. & Stewart, Ian B. (2008) Calibration of an ingestible temperature sensor. Physiological Measurement, 29(11), N71-N78.
Abstract
An ingestible telemetric sensor for measuring core body temperature is increasingly being utilized in occupational and athletic studies of heat strain. There is a need for a uniform method of calibrating these sensors in the scientific community in order to effectively compare the results of different researchers. The purpose of the present investigation was to determine and present such a calibration procedure. Sensors were placed in a water bath heated to nine discrete temperatures, and the recorded values were compared to that of a traceable thermometer. It was observed that sensor 2 recorded temperatures higher than sensors 1 and 3, and that all sensors were higher than the traceable thermometer, highlighting the need for a calibration procedure. The findings of this study suggest a number of recommendations for a calibration procedure including: (1) four water bath temperatures in the range of 33–41 degrees Celsius should be utilized; (2) sensors should be immersed for a minimum of 4 min prior to taking a measurement; (3) a linear regression relating sensor temperature to a traceable thermometer is an appropriate method to adjust raw data. Switching the sensor off after calibration and reactivating it prior to ingestion will not influence the accuracy of temperature measurement.
Citations:
Citation countsare sourced monthly from Scopus and Web of Science citation databases.
These databases contain citations from different subsets of available publications and different time periods and thus the citation count from each is usually different. Some works are not in either database and no count is displayed. Scopus includes citations from articles published in 1996 onwards, and Web of Science generally from 1980 onwards.
Citations counts from the Google Scholar™ indexing service can be viewed at the linked Google Scholar™ search.
| ID Code: | 15503 |
|---|---|
| Item Type: | Journal Article |
| Additional Information: | For more information, please refer to the journal's website (see hypertext link) or contact the author. |
| Keywords: | core body temperature, calibration, ingestible sensor |
| DOI: | 10.1088/0967-3334/29/11/N01 |
| ISSN: | 1361-6579 |
| Subjects: | Australian and New Zealand Standard Research Classification > MEDICAL AND HEALTH SCIENCES (110000) > HUMAN MOVEMENT AND SPORTS SCIENCE (110600) > Exercise Physiology (110602) |
| Divisions: | Current > QUT Faculties and Divisions > Faculty of Health Current > Institutes > Institute of Health and Biomedical Innovation |
| Copyright Owner: | Copyright 2008 IOP (Institute of Physics) Publishing |
| Deposited On: | 06 Nov 2008 |
| Last Modified: | 29 Feb 2012 23:46 |
Export: EndNote | Dublin Core | BibTeX
Repository Staff Only: item control page