Repeatability and reliability of ocular aberration measurements in contact lens wear
Efron, Suzanne, Efron, Nathan, & Morgan, Philip B. (2008) Repeatability and reliability of ocular aberration measurements in contact lens wear. Contact Lens and Anterior Eye, 31(2), pp. 81-88.
DOI: 10.1016/j.clae.2007.10.004
Official URL: http://nl.sitestat.com/elsevier/elsevier-com/s?Sci...
Abstract
Background
To assess repeatability and within- and between-observer reliability when measuring ocular aberrations with and without soft contact lens wear.
Methods
The irx3 aberrometer was used to measure second to tenth radial order aberrations on 13 subjects with and without their habitual soft contact lenses. This was first performed by one observer, who took three separate measurements to determine repeatability. In addition, two observers each took a set of three measurements on two occasions, to determine within- and between-observer reliability. Data were normalized to a constant pupil size for all measurements made on each subject.
Results
There was no difference in repeatability with or without contact lenses for Zernike polynomial numbers 3–20. There was no difference between within- and between-observer relative reliability either with lenses (t = 1.9, p = 0.07) or without lenses (t = 1.1, p = 0.29). However, both within-observer and between-observer reliability were worse when lenses were worn (t = 3.6, p = 0.0009 and t = 2.8, p = 0.0074, respectively).
Conclusion
The reduction in within- and between-observer reliability when measuring aberrations in contact lens wearers may be due to variability resulting from a variety of factors, such as lens flexure, lens movement, lens decentration and reduced stability of the pre-lens tear film.
Citations:

Citation counts are sourced monthly from Scopus and Web of Science citation databases.
These databases contain citations from different subsets of available publications and different time periods and thus the citation count from each is usually different. Some works are not in either database and no count is displayed. Scopus includes citations from articles published in 1996 onwards, and Web of Science generally from 1980 onwards.
Citations counts from the Google Scholar™ indexing service can be viewed at the linked Google Scholar™ search.
| ID Code: | 19022 |
|---|---|
| Item Type: | Journal Article |
| Keywords: | Aberration, Contact lens, Eye, Reliability, Repeatability |
| ISSN: | 13670484 |
| Subjects: | Australian and New Zealand Standard Research Classification > MEDICAL AND HEALTH SCIENCES (110000) > OPTOMETRY AND OPHTHALMOLOGY (111300) Australian and New Zealand Standard Research Classification > MEDICAL AND HEALTH SCIENCES (110000) > OPTOMETRY AND OPHTHALMOLOGY (111300) > Optometry and Ophthalmology not elsewhere classified (111399) |
| Divisions: | Current > QUT Faculties and Divisions > Faculty of Health Current > Institutes > Institute of Health and Biomedical Innovation Current > Schools > School of Optometry & Vision Science |
| Deposited On: | 07 Apr 2009 09:47 |
| Last Modified: | 02 Feb 2012 20:41 |
Export: EndNote | Dublin Core | BibTeX
Staff only: HERDC collection form
Repository Staff Only: item control page