QUT ePrints

Magnitude and variability of loading on the osseointegrated implant of transfemoral amputees during walking

Lee, Winson C.C., Frossard, Laurent A., Hagberg, Kerstin , Haggstrom, Eva , Lee Gow , David , & Gray , Steven (2008) Magnitude and variability of loading on the osseointegrated implant of transfemoral amputees during walking. Medical Engineering & Physics, 30(7), pp. 825-833.

View at publisher

Abstract

This study directly measured the load acting on the abutment of the osseointegrated implant system of transfemoral amputees during level walking, and studied the variability of the load within and among amputees. Twelve active transfemoral amputees (age: 54±12 years, mass:84.3±16.3 kg, height: 17.8±0.10 m) fitted with an osseointegrated implant for over 1 year participated in the study. The load applied on the abutment was measured during unimpeded, level walking in a straight line using a commercial six-channel transducer mounted between the abutment and the prosthetic knee. The pattern and the magnitude of the three-dimensional forces and moments were revealed. Results showed a low step-to-step variability of each subject, but a high subject-to-subject variability in local extrema of body-weight normalized forces and moments and impulse data. The high subject-to-subject variability suggests that the mechanical design of the implant system should be customized for each individual, or that a fit-all design should take into consideration the highest values of load within a broad range of amputees. It also suggests specific loading regime in rehabilitation training are necessary for a given subject. Thus the loading magnitude and variability demonstrated should be useful in designing an osseointegrated implant system better able to resist mechanical failure and in refining the rehabilitation protocol.

Impact and interest:

32 citations in Scopus
Search Google Scholar™
21 citations in Web of Science®

Citation countsare sourced monthly from Scopus and Web of Science® citation databases.

These databases contain citations from different subsets of available publications and different time periods and thus the citation count from each is usually different. Some works are not in either database and no count is displayed. Scopus includes citations from articles published in 1996 onwards, and Web of Science® generally from 1980 onwards.

Citations counts from the Google Scholar™ indexing service can be viewed at the linked Google Scholar™ search.

Full-text downloads:

331 since deposited on 07 Dec 2009
62 in the past twelve months

Full-text downloadsdisplays the total number of times this work’s files (e.g., a PDF) have been downloaded from QUT ePrints as well as the number of downloads in the previous 365 days. The count includes downloads for all files if a work has more than one.

ID Code: 29105
Item Type: Journal Article
Keywords: Transfemoral amputation, Prosthetics, Osseointegration, Transducer, Gait
DOI: 10.1016/j.medengphy.2007.09.003
ISSN: 1350-4533
Subjects: Australian and New Zealand Standard Research Classification > MEDICAL AND HEALTH SCIENCES (110000) > HUMAN MOVEMENT AND SPORTS SCIENCE (110600) > Human Movement and Sports Science not elsewhere classified (110699)
Divisions: Current > Institutes > Institute of Health and Biomedical Innovation
Copyright Owner: Copyright 2007 IPEM Published by Elsevier Ltd.
Deposited On: 07 Dec 2009 11:36
Last Modified: 11 Aug 2011 00:27

Export: EndNote | Dublin Core | BibTeX