Repeatability and validity of lens densitometry measured with Scheimpflug imaging
Kirkwood, Bradley J, Hendicott, Peter Leslie, Read, Scott A., & Pesudovs, Konrad (2009) Repeatability and validity of lens densitometry measured with Scheimpflug imaging. Journal of Cataract & Refractive Surgery, 35(7), pp. 1210-1215.
Abstract
Purpose
To assess the repeatability and validity of lens densitometry derived from the Pentacam Scheimpflug imaging system.
Setting
Eye Clinic, Queensland University of Technology, Brisbane, Australia.
Methods
This prospective cross-sectional study evaluated 1 eye of subjects with or without cataract. Scheimpflug measurements and slitlamp and retroillumination photographs were taken through a dilated pupil. Lenses were graded with the Lens Opacities Classification System III. Intraobserver and interobserver reliability of 3 observers performing 3 repeated Scheimpflug lens densitometry measurements each was assessed. Three lens densitometry metrics were evaluated: linear, for which a line was drawn through the visual axis and a mean lens densitometry value given; peak, which is the point at which lens densitometry is greatest on the densitogram; 3-dimensional (3D), in which a fixed, circular 3.0 mm area of the lens is selected and a mean lens densitometry value given. Bland and Altman analysis of repeatability for multiple measures was applied; results were reported as the repeatability coefficient and relative repeatability (RR).
Results
Twenty eyes were evaluated. Repeatability was high. Overall, interobserver repeatability was marginally lower than intraobserver repeatability. The peak was the least reliable metric (RR 37.31%) and 3D, the most reliable (RR 5.88%). Intraobserver and interobserver lens densitometry values in the cataract group were slightly less repeatable than in the noncataract group.
Conclusion
The intraobserver and interobserver repeatability of Scheimpflug lens densitometry was high in eyes with cataract and eyes without cataract, which supports the use of automated lens density scoring using the Scheimpflug system evaluated in the study
Citations:
Citation countsare sourced monthly from Scopus and Web of Science citation databases.
These databases contain citations from different subsets of available publications and different time periods and thus the citation count from each is usually different. Some works are not in either database and no count is displayed. Scopus includes citations from articles published in 1996 onwards, and Web of Science generally from 1980 onwards.
Citations counts from the Google Scholar™ indexing service can be viewed at the linked Google Scholar™ search.
Full-text downloads:
Full-text downloadsdisplays the total number of times this work’s files (e.g., a PDF) have been downloaded from QUT ePrints as well as the number of downloads in the previous 365 days. The count includes downloads for all files if a work has more than one.
| ID Code: | 29571 |
|---|---|
| Item Type: | Journal Article |
| DOI: | 10.1016/j.jcrs.2009.03.017 |
| ISSN: | 0886-3350 |
| Divisions: | Current > QUT Faculties and Divisions > Faculty of Health Current > Institutes > Institute of Health and Biomedical Innovation Current > Schools > School of Optometry & Vision Science |
| Copyright Owner: | Copyright 2009 Elsevier. |
| Deposited On: | 14 Jan 2010 09:08 |
| Last Modified: | 22 Feb 2013 11:55 |
Export: EndNote | Dublin Core | BibTeX