Typical accuracy and quality control of a process for creating CT-Based virtual bone models

Noser, Hansrudi, Heldstab, Thomas, Schmutz, Beat, & Kamer, Lukas (2010) Typical accuracy and quality control of a process for creating CT-Based virtual bone models. Journal of Digital Imaging.

View at publisher

Abstract

A pragmatic method for assessing the accuracy and precision of a given processing pipeline required for converting computed tomography (CT) image data of bones into representative three dimensional (3D) models of bone shapes is proposed. The method is based on coprocessing a control object with known geometry which enables the assessment of the quality of resulting 3D models. At three stages of the conversion process, distance measurements were obtained and statistically evaluated. For this study, 31 CT datasets were processed. The final 3D model of the control object contained an average deviation from reference values of −1.07±0.52 mm standard deviation (SD) for edge distances and −0.647±0.43 mm SD for parallel side distances of the control object. Coprocessing a reference object enables the assessment of the accuracy and precision of a given processing pipeline for creating CTbased 3D bone models and is suitable for detecting most systematic or human errors when processing a CT-scan. Typical errors have about the same size as the scan resolution.

Impact and interest:

5 citations in Scopus
Search Google Scholar™
5 citations in Web of Science®

Citation countsare sourced monthly from Scopus and Web of Science® citation databases.

These databases contain citations from different subsets of available publications and different time periods and thus the citation count from each is usually different. Some works are not in either database and no count is displayed. Scopus includes citations from articles published in 1996 onwards, and Web of Science® generally from 1980 onwards.

Citations counts from the Google Scholar™ indexing service can be viewed at the linked Google Scholar™ search.

Full-text downloads:

213 since deposited on 13 Aug 2010
81 in the past twelve months

Full-text downloadsdisplays the total number of times this work’s files (e.g., a PDF) have been downloaded from QUT ePrints as well as the number of downloads in the previous 365 days. The count includes downloads for all files if a work has more than one.

ID Code: 33727
Item Type: Journal Article
Keywords: Computed tomography, virtual bone models, size accuracy
DOI: 10.1007/s10278-010-9287-4
ISSN: 0897-1889
Subjects: Australian and New Zealand Standard Research Classification > MEDICAL AND HEALTH SCIENCES (110000) > CLINICAL SCIENCES (110300)
Divisions: Past > QUT Faculties & Divisions > Faculty of Built Environment and Engineering
Past > Schools > School of Engineering Systems
Copyright Owner: Copyright 2010 Springer
Copyright Statement: The original publication is available at SpringerLink http://www.springerlink.com
Deposited On: 13 Aug 2010 11:42
Last Modified: 11 Aug 2011 02:33

Export: EndNote | Dublin Core | BibTeX

Repository Staff Only: item control page