Systematic variations in polymer gel dosimeter calibration due to container influence and deviations from water equivalence

Taylor, M.L. , Franich, R.D. , Johnston, P.N. , Millar, R.M. , & Trapp, J.V. (2007) Systematic variations in polymer gel dosimeter calibration due to container influence and deviations from water equivalence. Physics in Medicine and Biology, 52(13), pp. 3991-4005.

View at publisher

Abstract

There are a number of gel dosimeter calibration methods in contemporary usage. The present study is a detailed Monte Carlo investigation into the accuracy of several calibration techniques. Results show that for most arrangements the dose to gel accurately reflects the dose to water, with the most accurate method involving the use of a large diameter flask of gel into which multiple small fields of varying dose are directed. The least accurate method was found to be that of a long test tube in a water phantom, coaxial with the beam. The large flask method is also the most straightforward and least likely to introduce errors during setup, though, to its detriment, the volume of gel required is much more than other methods.

Impact and interest:

16 citations in Scopus
Search Google Scholar™
13 citations in Web of Science®

Citation countsare sourced monthly from Scopus and Web of Science® citation databases.

These databases contain citations from different subsets of available publications and different time periods and thus the citation count from each is usually different. Some works are not in either database and no count is displayed. Scopus includes citations from articles published in 1996 onwards, and Web of Science® generally from 1980 onwards.

Citations counts from the Google Scholar™ indexing service can be viewed at the linked Google Scholar™ search.

Full-text downloads:

48 since deposited on 05 Oct 2010
18 in the past twelve months

Full-text downloadsdisplays the total number of times this work’s files (e.g., a PDF) have been downloaded from QUT ePrints as well as the number of downloads in the previous 365 days. The count includes downloads for all files if a work has more than one.

ID Code: 37727
Item Type: Journal Article
Additional URLs:
Keywords: gel dosimeter calibration, water phantom
DOI: 10.1088/0031-9155/52/13/022
ISSN: 0031-9155
Subjects: Australian and New Zealand Standard Research Classification > PHYSICAL SCIENCES (020000) > OTHER PHYSICAL SCIENCES (029900) > Medical Physics (029903)
Divisions: Past > QUT Faculties & Divisions > Faculty of Science and Technology
Past > Schools > Physics
Copyright Owner: Copyright 2007 Institute of Physics
Deposited On: 06 Oct 2010 08:13
Last Modified: 29 Feb 2012 23:52

Export: EndNote | Dublin Core | BibTeX

Repository Staff Only: item control page