A compact mock circulation loop for the in vitro testing of cardiovascular devices
Timms, Daniel L., Gregory, Shaun D., Greatrex, Nicholas A., Pearcy, Mark J., Fraser, John F., & Steinseifer, Ulrich (2010) A compact mock circulation loop for the in vitro testing of cardiovascular devices. Artificial Organs.
Abstract
In vitro cardiovascular device performance evaluation in a mock circulation loop (MCL) is a necessary step prior to in vivo testing.A MCL that accurately represents the physiology of the cardiovascular system accelerates
the assessment of the device’s ability to treat
pathological conditions. To serve this purpose, a compact
MCL measuring 600 ¥ 600 ¥ 600 mm (L ¥ W¥ H) was constructed
in conjunction with a computer mathematical simulation.This approach allowed the effective selection of physical loop characteristics, such as pneumatic drive parameters, to create pressure and flow, and pipe dimensions to replicate the resistance, compliance, and fluid inertia of the native cardiovascular system. The resulting five-element MCL reproduced the physiological hemodynamics of a healthy and failing heart by altering ventricle contractility, vascular resistance/compliance, heart rate, and vascular volume. The effects of interpatient anatomical variability, such as septal defects and valvular disease, were also assessed. Cardiovascular hemodynamic pressures (arterial, venous, atrial, ventricular), flows (systemic, bronchial, pulmonary), and volumes (ventricular, stroke)
were analyzed in real time. The objective of this study
is to describe the developmental stages of the compact
MCL and demonstrate its value as a research tool for
the accelerated development of cardiovascular devices.
Citations:
Citation countsare sourced monthly from Scopus and Web of Science citation databases.
These databases contain citations from different subsets of available publications and different time periods and thus the citation count from each is usually different. Some works are not in either database and no count is displayed. Scopus includes citations from articles published in 1996 onwards, and Web of Science generally from 1980 onwards.
Citations counts from the Google Scholar™ indexing service can be viewed at the linked Google Scholar™ search.
| ID Code: | 39877 |
|---|---|
| Item Type: | Journal Article |
| Keywords: | Mock Circulation Loop, Heart Disease, Cardiovascular System, Device Testing |
| DOI: | 10.1111/j.1525-1594.2010.01088.x |
| ISSN: | 0160564X |
| Subjects: | Australian and New Zealand Standard Research Classification > ENGINEERING (090000) > BIOMEDICAL ENGINEERING (090300) |
| Divisions: | Past > QUT Faculties & Divisions > Faculty of Built Environment and Engineering Current > Institutes > Institute of Health and Biomedical Innovation Past > Schools > School of Engineering Systems |
| Deposited On: | 03 Feb 2011 08:30 |
| Last Modified: | 01 Mar 2012 00:22 |
Export: EndNote | Dublin Core | BibTeX
Repository Staff Only: item control page