Direct measurement of 3D force and moment on lower-limb osseointegrated fixation

Lee, Winson C.C., Frossard, Laurent A., Hagberg, Kerstin, Haggstrom, Eva, Gow, David A., Gray, Stephen, & Branemark, Rickard (2006) Direct measurement of 3D force and moment on lower-limb osseointegrated fixation. In Symposium on the 3-D Analysis of Human Movement, 28-30 June 2006, Valenciennes, France.

Abstract

Implant loosening and mechanical failure are two major concerns for trans-femoral amputees using osseointegrated fixation. Understanding the true load applied on osseointegrated fixation is one important step in solving the two problems. Conventionally, the load magnitude can be calculated based on the motion of the prosthesis and the ground reaction forces measured by a force plate using inverse dynamics. However, this approach allowed only one or two walking steps to be measured and errors are compounded when involving more than one joint. In this study, we directly measured load acting on the abutment of an osseointegrated prosthesis with a load transducer for twelve trans-femoral amputees. Three-dimensional force and moment patterns at the abutment along a gait cycle were displayed. Step-to-step variability of each subject in various kinetic and temporal parameters as well as subject-to-subject variability in temporal parameters were found low. High subject-to-subject variability in force and moment data was demonstrated.

Impact and interest:

Citation countsare sourced monthly from Scopus and Web of Science® citation databases.

These databases contain citations from different subsets of available publications and different time periods and thus the citation count from each is usually different. Some works are not in either database and no count is displayed. Scopus includes citations from articles published in 1996 onwards, and Web of Science® generally from 1980 onwards.

Citations counts from the Google Scholar™ indexing service can be viewed at the linked Google Scholar™ search.

Full-text downloads:

477 since deposited on 18 May 2006
73 in the past twelve months

Full-text downloadsdisplays the total number of times this work’s files (e.g., a PDF) have been downloaded from QUT ePrints as well as the number of downloads in the previous 365 days. The count includes downloads for all files if a work has more than one.

ID Code: 4041
Item Type: Conference Paper
Additional URLs:
Keywords: osseointegration, prosthetics, transducer, transfemoral amputees, variability
Subjects: Australian and New Zealand Standard Research Classification > ENGINEERING (090000) > BIOMEDICAL ENGINEERING (090300) > Rehabilitation Engineering (090305)
Divisions: Past > QUT Faculties & Divisions > Faculty of Built Environment and Engineering
Copyright Owner: Copyright 2006 (please consult author)