QUT ePrints

Commissioning of a Geant4 based treatment plan simulation tool : linac model and dicom-rt interface

Cornelius, Iwan, Poole, Christopher, Langton, Christian M., Hill, Brendan, Middlebrook, Nigel, & Oborn, Brad (2010) Commissioning of a Geant4 based treatment plan simulation tool : linac model and dicom-rt interface. In 29th European Society for Therapeutic Radiology and Oncology Conference (ESTRO 29), 12-16 September 2010, Barcelona. (Unpublished)

View at publisher

Abstract

A Geant4 based simulation tool has been developed to perform Monte Carlo modelling of a 6 MV VarianTM iX clinac. The computer aided design interface of Geant4 was used to accurately model the LINAC components, including the Millenium multi-leaf collimators (MLCs). The simulation tool was verified via simulation of standard commissioning dosimetry data acquired with an ionisation chamber in a water phantom. Verification of the MLC model was achieved by simulation of leaf leakage measurements performed using GafchromicTM film in a solid water phantom. An absolute dose calibration capability was added by including a virtual monitor chamber into the simulation. Furthermore, a DICOM-RT interface was integrated with the application to allow the simulation of treatment plans in radiotherapy. The ability of the simulation tool to accurately model leaf movements and doses at each control point was verified by simulation of a widely used intensity-modulated radiation therapy (IMRT) quality assurance (QA) technique, the chair test.

Impact and interest:

Citation countsare sourced monthly from Scopus and Web of Science® citation databases.

These databases contain citations from different subsets of available publications and different time periods and thus the citation count from each is usually different. Some works are not in either database and no count is displayed. Scopus includes citations from articles published in 1996 onwards, and Web of Science® generally from 1980 onwards.

Citations counts from the Google Scholar™ indexing service can be viewed at the linked Google Scholar™ search.

Full-text downloads:

389 since deposited on 04 May 2011
98 in the past twelve months

Full-text downloadsdisplays the total number of times this work’s files (e.g., a PDF) have been downloaded from QUT ePrints as well as the number of downloads in the previous 365 days. The count includes downloads for all files if a work has more than one.

ID Code: 41556
Item Type: Conference Paper
Additional Information: Also presented at 1st Geant4 Australian School and User Workshop, NSW 11-16 April, 2011 University of Wollongong,
Additional URLs:
Keywords: geant4, radiotherapy, chair test, quality assurance, imrt, monte carlo, rapidarc
Subjects: Australian and New Zealand Standard Research Classification > PHYSICAL SCIENCES (020000) > OTHER PHYSICAL SCIENCES (029900) > Medical Physics (029903)
Divisions: Past > QUT Faculties & Divisions > Faculty of Science and Technology
Current > Institutes > Institute of Health and Biomedical Innovation
Past > Schools > Physics
Copyright Owner: Copyright 2011 [please consult the author]
Deposited On: 05 May 2011 09:18
Last Modified: 07 Sep 2013 17:55

Export: EndNote | Dublin Core | BibTeX

Repository Staff Only: item control page