Monte Carlo-based diode design for correction-less small field dosimetry

Charles, P.H., Crowe, S.B., Kairn, T., Knight, R.T., Hill, B., Kenny, J., Langton, C.M., & Trapp, J.V. (2013) Monte Carlo-based diode design for correction-less small field dosimetry. Physics in Medicine and Biology, 58(13), pp. 4501-4512.

View at publisher

Abstract

Due to their small collecting volume diodes are commonly used in small field dosimetry. However the relative sensitivity of a diode increases with decreasing small field size. Conversely, small air gaps have been shown to cause a significant decrease in the sensitivity of a detector as the field size is decreased. Therefore this study uses Monte Carlo simulations to look at introducing air upstream to diodes such that they measure with a constant sensitivity across all field sizes in small field dosimetry.

Varying thicknesses of air were introduced onto the upstream end of two commercial diodes (PTW 60016 photon diode and PTW 60017 electron diode), as well as a theoretical unenclosed silicon chip using field sizes as small as 5 mm × 5 mm . The metric D_(w,Q)/D_(Det,Q) used in this study represents the ratio of the dose to a point of water to the dose to the diode active volume, for a particular field size and location. The optimal thickness of air required to provide a constant sensitivity across all small field sizes was found by plotting D_(w,Q)/D_(Det,Q) as a function of introduced air gap size for various field sizes, and finding the intersection point of these plots. That is, the point at which D_(w,Q)/D_(Det,Q) was constant for all field sizes was found.

The optimal thickness of air was calculated to be 3.3 mm, 1.15 mm and 0.10 mm for the photon diode, electron diode and unenclosed silicon chip respectively. The variation in these results was due to the different design of each detector. When calculated with the new diode design incorporating the upstream air gap, k_(Q_clin 〖,Q〗_msr)^(f_clin 〖,f〗_msr ) was equal to unity to within statistical uncertainty (0.5 %) for all three diodes. Cross-axis profile measurements were also improved with the new detector design.

The upstream air gap could be implanted on the commercial diodes via a cap consisting of the air cavity surrounded by water equivalent material. The results for the unclosed silicon chip show that an ideal small field dosimetry diode could be created by using a silicon chip with a small amount of air above it.

Impact and interest:

25 citations in Scopus
24 citations in Web of Science®
Search Google Scholar™

Citation counts are sourced monthly from Scopus and Web of Science® citation databases.

These databases contain citations from different subsets of available publications and different time periods and thus the citation count from each is usually different. Some works are not in either database and no count is displayed. Scopus includes citations from articles published in 1996 onwards, and Web of Science® generally from 1980 onwards.

Citations counts from the Google Scholar™ indexing service can be viewed at the linked Google Scholar™ search.

Full-text downloads:

206 since deposited on 03 Jul 2013
28 in the past twelve months

Full-text downloads displays the total number of times this work’s files (e.g., a PDF) have been downloaded from QUT ePrints as well as the number of downloads in the previous 365 days. The count includes downloads for all files if a work has more than one.

ID Code: 61061
Item Type: Journal Article
Refereed: Yes
Keywords: radiotherapy, dosimetry, small field, monte carlo
DOI: 10.1088/0031-9155/58/13/4501
ISSN: 0031-9155
Subjects: Australian and New Zealand Standard Research Classification > PHYSICAL SCIENCES (020000) > OTHER PHYSICAL SCIENCES (029900) > Medical Physics (029903)
Divisions: Past > QUT Faculties & Divisions > Faculty of Science and Technology
Current > QUT Faculties and Divisions > Science & Engineering Faculty
Funding:
Copyright Owner: Copyright 2013 Institute of Physics
Deposited On: 03 Jul 2013 01:19
Last Modified: 30 Jun 2017 06:40

Export: EndNote | Dublin Core | BibTeX

Repository Staff Only: item control page