Verification of patient position and delivery of IMRT by electronic Portal Imaging

Fielding, A L, Clark, C H, & Evans, P M (2003) Verification of patient position and delivery of IMRT by electronic Portal Imaging. Medical Physics, 30(6), p. 1352.

[img] conference abstracts pdf page 25 (PDF 3MB)
Published Version.
Administrators only | Request a copy from author

View at publisher

Abstract

We have taken a new method of calibrating portal images of IMRT beams and used this to measure patient set-up accuracy and delivery errors, such as leaf errors and segment intensity errors during treatment. A calibration technique was used to remove the intensity modulations from the images leaving equivalent open field images that show patient anatomy that can be used for verification of the patient position. The images of the treatment beam can also be used to verify the delivery of the beam in terms of multileaf collimator leaf position and dosimetric errors. A series of controlled experiments delivering an IMRT anterior beam to the head and neck of a humanoid phantom were undertaken. A 2mm translation in the position of the phantom could be detected. With intentional introduction of delivery errors into the beam this method allowed us to detect leaf positioning errors of 2mm and variation in monitor units of 1%. The method was then applied to the case of a patient who received IMRT treatment to the larynx and cervical nodes. The anterior IMRT beam was imaged during four fractions and the images calibrated and investigated for the characteristic signs of patient position error and delivery error that were shown in the control experiments. No significant errors were seen. The method of imaging the IMRT beam and calibrating the images to remove the intensity modulations can be a useful tool in verifying both the patient position and the delivery of the beam.

Impact and interest:

Search Google Scholar™

Citation counts are sourced monthly from Scopus and Web of Science® citation databases.

These databases contain citations from different subsets of available publications and different time periods and thus the citation count from each is usually different. Some works are not in either database and no count is displayed. Scopus includes citations from articles published in 1996 onwards, and Web of Science® generally from 1980 onwards.

Citations counts from the Google Scholar™ indexing service can be viewed at the linked Google Scholar™ search.

ID Code: 61260
Item Type: Journal Article
Refereed: Yes
Additional URLs:
DOI: 10.1118/1.1595296
ISSN: 0094-2405 (print)
Divisions: Current > Institutes > Institute of Health and Biomedical Innovation
Current > QUT Faculties and Divisions > Science & Engineering Faculty
Copyright Owner: Copyright 2003 American Association of Physicists in Medicine
Deposited On: 15 Jul 2013 01:27
Last Modified: 24 Jun 2017 14:36

Export: EndNote | Dublin Core | BibTeX

Repository Staff Only: item control page