QUT ePrints

Re-design of the Exeter V40 long stem femoral component for ease of removal

Tevelen, Gregory A., Pearcy, Mark J., & Crawford, Ross W. (2007) Re-design of the Exeter V40 long stem femoral component for ease of removal. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 221(2), pp. 195-201.

View at publisher

Abstract

Clinical experience shows that removal of the Exeter long-stem femoral component (220 mm, 240 mm, 260 mm) of total hip arthroplasty is extremely difficult, often requiring splitting of the femur. To identify the reason for this, measurements of stem geometry and force required to pull the stems out of the cement mantle were conducted on three original Exeter long-stem and one standard femoral components. All implants required an initial force of approximately 4 kN for release from the cement. The long-stem components then required much larger forces and hence much higher expenditure of energy to pull them clear of the cement. This was attributed to the reverse taper seen on the nominally cylindrical distal section of the long-stem components. Following re-design of the manufacturing process to ensure the taper continued to the implant's distal tip, four further implants were tested. These demonstrated the requirement for initial cement release but then required no further energy expenditure similar to the standard stem. This study clearly demonstrated that the original difficulty in removing these long stems was owing to the manufacturing process resulting in a reverse taper on the distal stem. The adoption of recommended manufacturing changes to ensure the taper continues to the distal tip removed this difficulty.

Impact and interest:

0 citations in Scopus
Search Google Scholar™
0 citations in Web of Science®

Citation countsare sourced monthly from Scopus and Web of Science® citation databases.

These databases contain citations from different subsets of available publications and different time periods and thus the citation count from each is usually different. Some works are not in either database and no count is displayed. Scopus includes citations from articles published in 1996 onwards, and Web of Science® generally from 1980 onwards.

Citations counts from the Google Scholar™ indexing service can be viewed at the linked Google Scholar™ search.

ID Code: 8255
Item Type: Journal Article
Additional Information: For more information, please refer to the journal's website (see hypertext link) or contact the author. Author contact details: m.pearcy@qut.edu.au
Keywords: Exeter long stem, total hip replacement, revision
DOI: 10.1243/09544119JEIM170
ISSN: 0954-4119
Subjects: Australian and New Zealand Standard Research Classification > ENGINEERING (090000) > BIOMEDICAL ENGINEERING (090300) > Biomechanical Engineering (090302)
Divisions: Past > QUT Faculties & Divisions > Faculty of Built Environment and Engineering
Current > Institutes > Institute of Health and Biomedical Innovation
Copyright Owner: Copyright 2007 Professional Engineering Publishing
Deposited On: 26 Jun 2007
Last Modified: 29 Feb 2012 23:35

Export: EndNote | Dublin Core | BibTeX