Field size consistency of nominally matched linacs

Kairn, Tanya, Asena, Andre, Charles, Paul, Hill, Brendan, Langton, Christian M., Middlebrook, Nigel, Moylan, Rebecca, & Trapp, Jamie (2015) Field size consistency of nominally matched linacs. Australasian Physical and Engineering Sciences in Medicine, 38(2), pp. 289-297.

View at publisher

Abstract

Given that there is increasing recognition of the effect that submillimetre changes in collimator position can have on radiotherapy beam dosimetry, this study aimed to evaluate the potential variability in small field collimation that may exist between otherwise matched linacs. Field sizes and field output factors were measured using radiochromic film and an electron diode, for jaw- and MLC-collimated fields produced by eight dosimetrically matched Varian iX linacs (Varian Medical Systems, Palo Alto, USA). This study used nominal sizes from 0.6×0.6 to 10×10 cm215 , for jaw-collimated fields,and from 1×1 to 10×10 cm216 , for MLC-collimated fields, delivered from a zero (head up, beam directed vertically downward) gantry angle. Differences between the field sizes measured for the eight linacs exceeded the uncertainty of the film measurements and the repositioning uncertainty of the jaws and MLCs on one linac. The dimensions of fields defined by MLC leaves were more consistent between linacs, while also differing more from their nominal values than fields defined by orthogonal jaws. The field output factors measured for the different linacs generally increased with increasing measured field size for the nominal 0.6×0.6 and 1×1 cm2 fields, and became consistent between linacs for nominal field sizes of 2×2 cm2 25 and larger. The inclusion in radiotherapy treatment planning system beam data of small field output factors acquired in fields collimated by jaws (rather than the more-reproducible MLCs), associated with either the nominal or the measured field sizes, should be viewed with caution. The size and reproducibility of the fields (especially the small fields) used to acquire treatment planning data should be investigated thoroughly as part of the linac or planning system commissioning process. Further investigation of these issues, using different linac models, collimation systems and beam orientations, is recommended.

Impact and interest:

1 citations in Scopus
Search Google Scholar™
1 citations in Web of Science®

Citation counts are sourced monthly from Scopus and Web of Science® citation databases.

These databases contain citations from different subsets of available publications and different time periods and thus the citation count from each is usually different. Some works are not in either database and no count is displayed. Scopus includes citations from articles published in 1996 onwards, and Web of Science® generally from 1980 onwards.

Citations counts from the Google Scholar™ indexing service can be viewed at the linked Google Scholar™ search.

Full-text downloads:

8 since deposited on 27 May 2015
6 in the past twelve months

Full-text downloads displays the total number of times this work’s files (e.g., a PDF) have been downloaded from QUT ePrints as well as the number of downloads in the previous 365 days. The count includes downloads for all files if a work has more than one.

ID Code: 84523
Item Type: Journal Article
Refereed: Yes
Keywords: linear accelerator, multi-leaf collimator, radiation therapy
DOI: 10.1007/s13246-015-0349-2
ISSN: 1879-5447
Subjects: Australian and New Zealand Standard Research Classification > PHYSICAL SCIENCES (020000) > OTHER PHYSICAL SCIENCES (029900) > Medical Physics (029903)
Divisions: Current > Schools > School of Chemistry, Physics & Mechanical Engineering
Current > QUT Faculties and Divisions > Science & Engineering Faculty
Funding:
Copyright Owner: Copyright 2015 Springer
Copyright Statement: The final publication is available at Springer via http://dx.doi.org/10.1007/s13246-015-0349-2
Deposited On: 27 May 2015 00:01
Last Modified: 02 Jul 2016 10:19

Export: EndNote | Dublin Core | BibTeX

Repository Staff Only: item control page