The effect of monochromatic aberrations on Autoref R-1 readings
Collins, Michael J. (2001) The effect of monochromatic aberrations on Autoref R-1 readings. Ophthalmic and Physiological Optics, 21(3), pp. 217-227.
Abstract
The Canon Autoref R-1 is used extensively to investigate aspects of the accommodation response of the human eye. The measurement zone of the Autoref R-1 is known to lie outside the central pupil and the instrument's readings are therefore potentially influenced by the monochromatic aberrations of the eye. Since these aberrations are known to change as a function of accommodation, the error associated with the aberrations might influence the stimulus-response slope. We used masks and contact lenses with known amounts of spherical aberration to establish the region of the pupil used by the Autoref R-1 in deriving a reading of the refractive power of the eye. This region was estimated to be narrowly centred on a 3.5-mm diameter. The spherical aberration of the eye was measured for five subjects across a range of accommodation levels and used to predict the likely effect upon Autoref R-1 readings. Positive spherical aberration will result in an apparent lead of accommodation, while negative spherical aberration causes an apparent lag of accommodation. For eyes with normal levels of spherical aberration, the slope of the stimulus-response will appear to be marginally flatter when measured with the Autoref R-1. An example of an eye with coma-like aberrations is also shown to produce Autoref R-1 results that will differ significantly from more central refractive measurements.
Citations:
Citation countsare sourced monthly from Scopus and Web of Science citation databases.
These databases contain citations from different subsets of available publications and different time periods and thus the citation count from each is usually different. Some works are not in either database and no count is displayed. Scopus includes citations from articles published in 1996 onwards, and Web of Science generally from 1980 onwards.
Citations counts from the Google Scholar™ indexing service can be viewed at the linked Google Scholar™ search.
Full-text downloads:
Full-text downloadsdisplays the total number of times this work’s files (e.g., a PDF) have been downloaded from QUT ePrints as well as the number of downloads in the previous 365 days. The count includes downloads for all files if a work has more than one.
| ID Code: | 881 |
|---|---|
| Item Type: | Journal Article |
| Additional Information: | http://www.hlth.qut.edu.au/opt/research/contacts.jsp Michael J. Collins (email):m.collins@qut.edu.au |
| DOI: | 10.1016/S0275-5408(00)00030-2 |
| ISSN: | 0275-5408 |
| Subjects: | Australian and New Zealand Standard Research Classification > MEDICAL AND HEALTH SCIENCES (110000) > OPTOMETRY AND OPHTHALMOLOGY (111300) |
| Divisions: | Current > QUT Faculties and Divisions > Faculty of Health Current > Institutes > Institute of Health and Biomedical Innovation |
| Copyright Owner: | Copyright 2001 Elsevier |
| Copyright Statement: | Reproduced in accordance with the copyright policy of the publisher. |
| Deposited On: | 21 Mar 2005 |
| Last Modified: | 09 Jun 2010 22:23 |