The effect of beam arrangements and the impact of non-coplanar beams on the treatment planning of stereotactic ablative radiation therapy for early stage lung cancer

Fitzgerald, Rhys, Owen, Rebecca, Barry, Tamara, Hargrave, Cathy, Pryor, David, Bernard, Anne, Lehman, Margot, Mai, Tao, & Fielding, Andrew (2016) The effect of beam arrangements and the impact of non-coplanar beams on the treatment planning of stereotactic ablative radiation therapy for early stage lung cancer. Journal of Medical Radiation Sciences, 63(1), pp. 31-40.

View at publisher (open access)



  • The aim of this study was to compare various coplanar and non-coplanar 3-dimensional conformal radiation therapy (3DCRT) beam arrangements for the delivery of stereotactic ablative radiation therapy (SABR) to patients with early stage lung cancer, based on the dosimetric criteria from the Radiation Therapy Oncology Group (RTOG) 1021 protocol.


  • Ten medically inoperable lung cancer patients eligible for SABR were re-planned using three different coplanar and three different non-coplanar beam arrangements. The plans were compared by assessing planning target volume (PTV) coverage, doses to normal tissues, the high-dose conformity (conformity index) and intermediate dose spillage as defined by the D2cm, (the dose at any point 2 cm away from the PTV), and the R50% (the ratio of the volume of half the prescription dose to the volume of the PTV).


  • Sixty plans in total were assessed. Mean PTV coverage with the prescription isodose was similar between coplanar (95.14%) and non-coplanar (95.26%) techniques (P = 0.47). There was significant difference between all coplanar and all non-coplanar fields for the R50% (P < 0.0001) but none for the D2cm (P = 0.19). The seven and nine field beam arrangements with two non-coplanar fields had less unacceptable protocol deviations (10 and 7) than the seven and nine field plans with only coplanar fields (13 and 8). The 13 field coplanar fields did not improve protocol compliance with eight unacceptable deviations. The 10 field non-coplanar beam arrangement achieved best compliance with the RTOG 1021 dose criteria with only one unacceptable deviation (maximum rib dose).


  • A 3DCRT planning technique using 10 fields with ≥6 non-coplanar beams best satisfied high and intermediate dose constraints stipulated in the RTOG 1021 trial. Further investigations are required to determine if minor protocol deviations should be balanced against efficiency with the extended treatment times required to deliver non-coplanar fields and if treatment times can be improved using novel intensity modulated techniques.

Impact and interest:

2 citations in Web of Science®
Search Google Scholar™

Citation counts are sourced monthly from Scopus and Web of Science® citation databases.

These databases contain citations from different subsets of available publications and different time periods and thus the citation count from each is usually different. Some works are not in either database and no count is displayed. Scopus includes citations from articles published in 1996 onwards, and Web of Science® generally from 1980 onwards.

Citations counts from the Google Scholar™ indexing service can be viewed at the linked Google Scholar™ search.

Full-text downloads:

1 since deposited on 13 Oct 2016
1 in the past twelve months

Full-text downloads displays the total number of times this work’s files (e.g., a PDF) have been downloaded from QUT ePrints as well as the number of downloads in the previous 365 days. The count includes downloads for all files if a work has more than one.

ID Code: 99949
Item Type: Journal Article
Refereed: Yes
DOI: 10.1002/jmrs.118
ISSN: 2051-3895
Divisions: Current > Schools > School of Chemistry, Physics & Mechanical Engineering
Current > Institutes > Institute of Health and Biomedical Innovation
Current > QUT Faculties and Divisions > Science & Engineering Faculty
Copyright Owner: 2015 The Authors. Journal of Medical Radiation Sciences published by Wiley Publishing Asia Pty Ltd on behalf of Australian Institute of Radiography and New Zealand Institute of Medical Radiation Technology.
Copyright Statement: This is an open access article under the terms of the Creative Commons Attribution-NonCommercial-NoDerivs License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non-commercial and no modifications or adaptations are made.