QUT QUT ePrints

A quasi-dynamic nonlinear finite element model to investigate prosthetic interface stresses during walking for trans-tibial amputees

Jia, Xiaohong and Zhang, Ming and Li, Xiaobing and Lee, Winson C. (2005) A quasi-dynamic nonlinear finite element model to investigate prosthetic interface stresses during walking for trans-tibial amputees. Clinical Biomechanics 20(6):pp. 630-635.

Full text available as:
PDF (Author Version) - Requires Adobe Acrobat Reader or other PDF viewer.

Abstract

Background. To predict the interface pressure between residual limb and prosthetic socket for trans-tibial amputees during walking. Methods. A quasi-dynamic finite element model was built based on the actual geometry of residual limb, internal bones and socket liner. To simulate the friction/slip boundary conditions between the skin and liner, automated surface-to-surface contact was used. Besides variable external loads and material inertia, the coupling between the large rigid displacement of knee joint and small elastic deformation of residual limb and prosthetic components were also considered. Results. Interface pressure distribution was found to have the same profile during walking. The high pressures fall over popliteal depression, middle patella tendon, lateral tibia and medial tibia regions. Interface pressure predicted by static or quasi-dynamic analysis had the similar double-peaked waveform shape in stance phase. Interpretation. The consideration of inertial effects and motion of knee joint cause 210% average variation of the area between the pressure curve and the horizontal line of pressure threshold between two cases, even though there is only a small change in the peak pressure. The findings in this paper show that the coupling dynamic effects of inertial loads and knee flexion must be considered to study interface pressure between residual limb and prosthetic socket during walking.

Item Type:Journal Article
Status:Published
Keywords:Prosthetic socket; Finite element analysis; Interface pressure; Knee flexion; Inertial loads
Subjects:290000 Engineering and Technology > 291500 Biomedical Engineering > 291502 Rehabilitation Engineering
ID Code:4341
Deposited By:Lee, Winson
Deposited On:29 June 2007
Alternative Locations:http://dx.doi.org/doi:10.1016/j.clinbiomech.2005.03.001
Copyright Owner:Copyright 2005 Elsevier
Copyright Statement:Reproduced in accordance with the copyright policy of the publisher.