Does forearm referencing using a retroversion guide achieve the targeted retroversion of the humeral component in reverse shoulder arthroplasty?

Jomaa, Mohammad N., , Ingoe, Helen, Pareyon, Roberto, Italia, Kristine, , Salhi, Asma, Gilliand, Luke, Nielsen, James, , , & (2024) Does forearm referencing using a retroversion guide achieve the targeted retroversion of the humeral component in reverse shoulder arthroplasty? Journal of Shoulder and Elbow Surgery, 33(7), pp. 1555-1562.

Open access copy at publisher website

Description

Background: Component positioning affects clinical outcomes of reverse shoulder arthroplasty, which necessitates an implantation technique that is reproducible, consistent, and reliable. This study aims to assess the accuracy and precision of positioning the humeral component in planned retroversion using a forearm referencing guide. Methods: Computed tomography scans of 54 patients (27 males and 27 females) who underwent primary reverse shoulder arthroplasty for osteoarthritis or cuff tear arthropathy were evaluated. A standardized surgical technique was used to place the humeral stem in 15° of retroversion. Version was assessed intraoperatively visualizing the retroversion guide from above and referencing the forearm axis. Metal subtraction techniques from postoperative computed tomography images allowed for the generation of 3D models of the humerus and for evaluation of the humeral component position. Anatomical humeral plane and implant planes were defined and the retroversion 3D angle between identified planes was recorded for each patient. Accuracy and precision were assessed. A subgroup analysis evaluated differences between male and female patients. Results: The humeral retroversion angle ranged from 0.9° to 22.8°. The majority (81%) of the measurements were less than 15°. Mean retroversion angle (±SD) was 9.9° ± 5.8° (95% CI 8.4°-11.5°) with a mean percent error with respect to 15° of −34% ± 38 (95% CI –23 to −44). In the male subgroup (n = 27, range 3.8°-22.5°), the mean retroversion angle was 11.9° ± 5.4° (95% CI 9.8°-14.1°) with a mean percent error with respect to 15° of −21% ± 36 (95% CI –6 to −35). In the female subgroup (n = 27, range 0.9°-22.8°), mean retroversion angle was 8.0° ± 5.5° (95% CI 5.8°-10.1°) and the mean percent error with respect to 15° was −47% ± 36 (95% CI –32 to −61). The differences between the 2 gender groups were statistically significant (P = .006). Conclusion: Referencing the forearm using an extramedullary forearm referencing system to position the humeral stem in a desired retroversion is neither accurate nor precise. There is a nonnegligible tendency to achieve a lower retroversion than planned, and the error is more marked in females.

Impact and interest:

Search Google Scholar™

Citation counts are sourced monthly from Scopus and Web of Science® citation databases.

These databases contain citations from different subsets of available publications and different time periods and thus the citation count from each is usually different. Some works are not in either database and no count is displayed. Scopus includes citations from articles published in 1996 onwards, and Web of Science® generally from 1980 onwards.

Citations counts from the Google Scholar™ indexing service can be viewed at the linked Google Scholar™ search.

ID Code: 250746
Item Type: Contribution to Journal (Journal Article)
Refereed: Yes
Measurements or Duration: 8 pages
Keywords: computer-generated 3D models, forearm-axis, humerus, Level II, retroversion, Reverse shoulder arthroplasty, Study of a Diagnostic Test
DOI: 10.1016/j.jse.2023.10.038
ISSN: 1058-2746
Pure ID: 173386925
Divisions: Current > QUT Faculties and Divisions > Faculty of Engineering
Current > Schools > School of Mechanical, Medical & Process Engineering
Copyright Owner: 2023 The Author(s)
Copyright Statement: This work is covered by copyright. Unless the document is being made available under a Creative Commons Licence, you must assume that re-use is limited to personal use and that permission from the copyright owner must be obtained for all other uses. If the document is available under a Creative Commons License (or other specified license) then refer to the Licence for details of permitted re-use. It is a condition of access that users recognise and abide by the legal requirements associated with these rights. If you believe that this work infringes copyright please provide details by email to qut.copyright@qut.edu.au
Deposited On: 18 Jul 2024 23:25
Last Modified: 19 Jul 2024 04:26